Test Catalog

Test Id : BVRNA

Bacterial Vaginosis, Nucleic Acid Amplification, Vaginal

Useful For
Suggests clinical disorders or settings where the test may be helpful

Aid for diagnosis of bacterial vaginosis

 

This test is not intended for use in medico-legal applications.

Method Name
A short description of the method used to perform the test

Transcription Mediated Amplification

NY State Available
Indicates the status of NY State approval and if the test is orderable for NY State clients.

Yes

Reporting Name
Lists a shorter or abbreviated version of the Published Name for a test

Bacterial Vaginosis, Amplified RNA

Aliases
Lists additional common names for a test, as an aid in searching

Bacterial Vaginosis

Vaginitis

Specimen Type
Describes the specimen type validated for testing

Vaginal

Specimen Required
Defines the optimal specimen required to perform the test and the preferred volume to complete testing

Specimen Type: Vaginal

Supplies: Aptima Multitest Swab Collection Kit (T584)

Container/Tube: Aptima Multitest Swab

Specimen Volume: Swab

Collection Instructions:

1. Specimen must be collected using the Aptima Multitest Swab Collection Kit.

2. Insert swab (pink shaft) about 5 cm past introitus and rotate gently for 30 seconds.

3. Place swab into transport tube provided in collection kit. Snap off swab at score line so swab fits into closed tube.

4. Cap tube securely, and label tube with patient's entire name and collection date and time.

5. Maintain swab container between 2 and 30 degrees C (refrigerate temperature is preferred) and transport within 30 days of collection. If longer storage is needed, store frozen between -20 and -70 degrees C for up to 60 days.

Specimen Minimum Volume
Defines the amount of sample necessary to provide a clinically relevant result as determined by the testing laboratory. The minimum volume is sufficient for one attempt at testing.

See Specimen Required

Reject Due To
Identifies specimen types and conditions that may cause the specimen to be rejected

Incorrect swab
Transport tube containing more than one swab
Reject

Specimen Stability Information
Provides a description of the temperatures required to transport a specimen to the performing laboratory, alternate acceptable temperatures are also included

Specimen Type Temperature Time Special Container
Vaginal Refrigerated (preferred) 30 days APTIMA VIAL
Frozen 60 days APTIMA VIAL
Ambient 30 days APTIMA VIAL

Useful For
Suggests clinical disorders or settings where the test may be helpful

Aid for diagnosis of bacterial vaginosis

 

This test is not intended for use in medico-legal applications.

Clinical Information
Discusses physiology, pathophysiology, and general clinical aspects, as they relate to a laboratory test

The Aptima BV (bacterial vaginosis) assay is intended to aid in the diagnosis of BV in individuals with clinical presentation consistent with vaginitis and/or vaginosis. Vaginitis is characterized by a spectrum of signs and symptoms, including vaginal/vulvar irritation, odor, discharge, and pruritus. Vaginitis may develop as a result of mechanical or chemical irritants (eg, feminine hygiene products, contraceptive materials) or due to a dysbiosis of the microbiota in the vaginal tract. Up to 90% of vaginitis cases are infectious, due to BV, vulvovaginal candidiasis (Candida vaginitis: CV) and/or trichomoniasis (Trichomonas vaginalis: TV). BV, CV, and TV individually account for 22% to 50%, 17% to 39%, and 4% to 35% of vaginitis cases, respectively. BV has been associated with pelvic inflammatory disease, cervicitis, elevated risk of acquisition of sexually transmitted infections (such as chlamydia, gonorrhea, herpes simplex virus, and HIV), spontaneous abortion, and preterm birth.

 

Bacterial vaginosis is characterized by a change in the vaginal microbiota dominated by Lactobacillus species to a polymicrobial anaerobe-dominated microbiota that includes Gardnerella vaginalis, Atopobium vaginae, Prevotella, Bacteroides, Peptostreptococcus, Mobiluncus, Sneathia (Leptotrichia), Mycoplasma, and BV-associated bacteria. A change in the normal vaginal microbiota is associated with the development of multiple signs and symptoms (eg, discharge, vaginal discomfort, and discharge). Diagnosis of BV can alternatively be established based on clinical criteria alone, referred to as Amsel's criteria, which include measuring vaginal pH, assessment for the presence of clue cells (eg, epithelial cells layered with bacterial cells), discharge, and malodor.

Reference Values
Describes reference intervals and additional information for interpretation of test results. May include intervals based on age and sex when appropriate. Intervals are Mayo-derived, unless otherwise designated. If an interpretive report is provided, the reference value field will state this.

Negative

Interpretation
Provides information to assist in interpretation of the test results

Positive: Results should be interpreted alongside clinical presentation. Up to 40% of asymptomatic patients may test positive by this assay. Assay result is based on relative amounts of Lactobacillus (Lactobacillus gasseri, Lactobacillus crispatus, Lactobacillus jensenii), Gardnerella vaginalis, and Atopobium vaginae. Individual organisms are not reported.

 

Negative: A negative result does not exclude infection. Assay result is based on relative amounts of Lactobacillus (L gasseri, L crispatus, L jensenii), G vaginalis, and A vaginae.

 

Inconclusive: Repeat testing on a new sample is recommended if clinically indicated.

Cautions
Discusses conditions that may cause diagnostic confusion, including improper specimen collection and handling, inappropriate test selection, and interfering substances

Therapeutic failure or success cannot be determined with the Aptima bacterial vaginosis (BV) assay since nucleic acid may persist following appropriate antimicrobial therapy.

 

Bacterial species targeted by the Aptima BV assay may comprise part of the normal microbiome for a significant number of women; a BV positive result should be interpreted in conjunction with other clinical data available to the clinician.

 

A negative result does not preclude a possible infection. Reliable results are dependent on adequate specimen collection, transport, storage, and processing.

 

Performance of the assay has not been evaluated in individuals younger than 14 years.

 

Additional microorganisms not detected by the Aptima BV assay such as Prevotella species, Mobiluncus species, Ureaplasma, Mycoplasma, and numerous fastidious or uncultivated anaerobes have also been found in women with BV but are less associated with BV due to their relatively low prevalence, sensitivity, and/or specificity.

 

Interference with the Aptima BV assay was observed in the presence of the following substances: mucus (1.5% V/V), vaginal moisturizing gel (0.5% W/V) and tioconazole (5% W/V).

 

Cross-reactivity was observed with the Aptima BV assay in the presence of Lactobacillus acidophilus (1x10(4) CFU/mL).

Clinical Reference
Recommendations for in-depth reading of a clinical nature

1. Workowski KA, Bachmann LH, Chan PA, et al. Sexually transmitted infections Treatment Guidelines, 2021. MMWR Recomm Rep. 2021;70(4):1-187

2. Muzny CA, Cerca N, Elnaggar JH, Taylor CM, Sobel JD, Van Der Pol B. State of the art for diagnosis of bacterial vaginosis. J Clin Microbiol. 2023;61(8):e0083722. doi:10.1128/jcm.00837-22

Method Description
Describes how the test is performed and provides a method-specific reference

The Aptima BV (bacterial vaginosis) assay is an in vitro nucleic acid amplification test that utilizes real time transcription-mediated amplification for detection and quantitation of ribosomal RNA from bacteria associated with BV, including Lactobacillus (Lactobacillus gasseri, Lactobacillus crispatus, Lactobacillus jensenii), Gardnerella vaginalis, and Atopobium vaginae. The Panther system detects and discriminates between four fluorescent signals corresponding to Lactobacillus group, Atopobium vaginae, Gardnerella vaginalis, and IC amplification products. Signal emergence times for each target organism are compared to calibration information to determine the BV Positive or Negative status of each sample.(Package Insert: Aptima BV Assay. Hologic, Inc; 9/2020)

PDF Report
Indicates whether the report includes an additional document with charts, images or other enriched information

No

Day(s) Performed
Outlines the days the test is performed. This field reflects the day that the sample must be in the testing laboratory to begin the testing process and includes any specimen preparation and processing time before the test is performed. Some tests are listed as continuously performed, which means that assays are performed multiple times during the day.

Monday through Sunday

Report Available
The interval of time (receipt of sample at Mayo Clinic Laboratories to results available) taking into account standard setup days and weekends. The first day is the time that it typically takes for a result to be available. The last day is the time it might take, accounting for any necessary repeated testing.

1 to 4 days

Specimen Retention Time
Outlines the length of time after testing that a specimen is kept in the laboratory before it is discarded

7 days

Performing Laboratory Location
Indicates the location of the laboratory that performs the test

Rochester

Fees
Several factors determine the fee charged to perform a test. Contact your U.S. or International Regional Manager for information about establishing a fee schedule or to learn more about resources to optimize test selection.

  • Authorized users can sign in to Test Prices for detailed fee information.
  • Clients without access to Test Prices can contact Customer Service 24 hours a day, seven days a week.
  • Prospective clients should contact their account representative. For assistance, contact Customer Service.

Test Classification
Provides information regarding the medical device classification for laboratory test kits and reagents. Tests may be classified as cleared or approved by the US Food and Drug Administration (FDA) and used per manufacturer instructions, or as products that do not undergo full FDA review and approval, and are then labeled as an Analyte Specific Reagent (ASR) product.

This test has been cleared, approved, or is exempt by the US Food and Drug Administration and is used per manufacturer's instructions. Performance characteristics were verified by Mayo Clinic in a manner consistent with CLIA requirements.

CPT Code Information
Provides guidance in determining the appropriate Current Procedural Terminology (CPT) code(s) information for each test or profile. The listed CPT codes reflect Mayo Clinic Laboratories interpretation of CPT coding requirements. It is the responsibility of each laboratory to determine correct CPT codes to use for billing.

CPT codes are provided by the performing laboratory.

81513

LOINC® Information
Provides guidance in determining the Logical Observation Identifiers Names and Codes (LOINC) values for the order and results codes of this test. LOINC values are provided by the performing laboratory.

Test Id Test Order Name Order LOINC Value
BVRNA Bacterial Vaginosis, Amplified RNA 103590-6
Result Id Test Result Name Result LOINC Value
Applies only to results expressed in units of measure originally reported by the performing laboratory. These values do not apply to results that are converted to other units of measure.
620738 Bacterial Vaginosis, Amplified RNA 103590-6

Test Setup Resources

Setup Files
Test setup information contains test file definition details to support order and result interfacing between Mayo Clinic Laboratories and your Laboratory Information System.

Excel | Pdf

Sample Reports
Normal and Abnormal sample reports are provided as references for report appearance.

Normal Reports | Abnormal Reports

SI Sample Reports
International System (SI) of Unit reports are provided for a limited number of tests. These reports are intended for international account use and are only available through MayoLINK accounts that have been defined to receive them.

SI Normal Reports | SI Abnormal Reports

Test Update Resources

Change Type Effective Date
New Test 2024-08-20