Test Catalog

Test Id : HSTNI

Troponin I, High Sensitivity, Plasma

Useful For
Suggests clinical disorders or settings where the test may be helpful

Excluding the diagnosis of acute coronary syndromes

 

Explaining troponin T elevations related to skeletal myopathy and/or assay interferences

Method Name
A short description of the method used to perform the test

Electrochemiluminescence Immunoassay

NY State Available
Indicates the status of NY State approval and if the test is orderable for NY State clients.

Yes

Reporting Name
Lists a shorter or abbreviated version of the Published Name for a test

Troponin I, High Sensitivity, P

Aliases
Lists additional common names for a test, as an aid in searching

TnI

hsTnI

Specimen Type
Describes the specimen type validated for testing

Plasma

Specimen Required
Defines the optimal specimen required to perform the test and the preferred volume to complete testing

Collection Container/Tube:

Preferred: Mint green top (lithium heparin gel)

Acceptable: Green top (lithium heparin)

Submission Container/Tube: Plastic vial

Specimen Volume: 1 mL

Collection Instructions:

1. Plasma gel tube should be centrifuged within 2 hours of collection.

2. Green-top tube should be centrifuged, and the plasma aliquoted into a plastic vial within 2 hours of collection.

Forms

If not ordering electronically, complete, print, and send a Cardiovascular Test Request Form (T724) with the specimen.

Specimen Minimum Volume
Defines the amount of sample necessary to provide a clinically relevant result as determined by the testing laboratory. The minimum volume is sufficient for one attempt at testing.

0.5 mL

Reject Due To
Identifies specimen types and conditions that may cause the specimen to be rejected

Gross hemolysis Reject

Specimen Stability Information
Provides a description of the temperatures required to transport a specimen to the performing laboratory, alternate acceptable temperatures are also included

Specimen Type Temperature Time Special Container
Plasma Frozen (preferred) 180 days
Refrigerated 48 hours
Ambient 4 hours

Useful For
Suggests clinical disorders or settings where the test may be helpful

Excluding the diagnosis of acute coronary syndromes

 

Explaining troponin T elevations related to skeletal myopathy and/or assay interferences

Clinical Information
Discusses physiology, pathophysiology, and general clinical aspects, as they relate to a laboratory test

Troponin is a complex that regulates the contraction of striated muscle. It consists of 3 subunits (C, T, and I) that are located periodically along the thin filament of the myofibrils. Troponin I inhibits actomyosin ATPase.

 

Troponin I is an inhibitory protein and exhibits in 3 isoforms: cardiac muscle, slow-twitch skeletal muscle, and fast-twitch skeletal muscle. The cardiac form of troponin I has 31 amino acid residues on its N-terminal, which allow for specific polyclonal and monoclonal antibody development, as they are not present in the skeletal forms. The cardiac specificity of this isoform improves the accuracy of diagnosis in patients with acute or chronic skeletal muscle injury and possible concomitant myocardial injury.

 

Troponin I is the only troponin isotope present in the myocardium and is not expressed during any developmental stage in skeletal muscle. Troponin I is released into the bloodstream within hours of the onset of symptoms of myocardial infarction or ischemic damage. It can be detected at 3 to 6 hours following onset of chest pain, with peak concentrations at 12 to 16 hours, and remains elevated for 5 to 9 days.

Reference Values
Describes reference intervals and additional information for interpretation of test results. May include intervals based on age and sex when appropriate. Intervals are Mayo-derived, unless otherwise designated. If an interpretive report is provided, the reference value field will state this.

Males > or =18 years: < or =20 ng/L

Females > or =18 years: < or =15 ng/L

 

Reference values have not been established for patients younger than 18 years old.

Interpretation
Provides information to assist in interpretation of the test results

Elevations in cardiac troponin T (cTnT) can be due to skeletal muscle disease and not cardiac disease in certain circumstances. One way to unmask such elevations is to measure cardiac troponin I (cTnI), which will be normal in that situation. In addition, at times there are interferences that can cause spurious increases or decreases in cTnT values. Conceptually, these same interferences can occur with cTnI but in any given case, the likelihood of having both assays be confounded in that way is highly unusual. Thus, potential false-positive results would be unmasked by a normal cTnI and false-negative results by an elevated value.

Cautions
Discusses conditions that may cause diagnostic confusion, including improper specimen collection and handling, inappropriate test selection, and interfering substances

A troponin value above the upper reference limit (99th percentile) value is not always indicative of myocardial infarction. Other conditions resulting in myocardial cell damage can contribute to elevated cardiac troponin I levels. These conditions include, but are not limited to, myocarditis, cardiac surgery, angina, unstable angina, congestive heart failure, and noncardiac-related causes, such as, kidney failure and pulmonary embolism.

Clinical Reference
Recommendations for in-depth reading of a clinical nature

Apple FS, Wu AHB, Sandoval Y, et al. Sex-specific 99th percentile upper reference limits for high sensitivity cardiac troponin assays derived using a universal sample bank. Clin Chem. 2020;66(3):434-444

Method Description
Describes how the test is performed and provides a method-specific reference

The Access hsTnI (high sensitivity troponin I) assay is a two-site immunoenzymatic (sandwich) assay. Monoclonal anti-cTnI (cardiac troponin I) antibody conjugated to alkaline phosphatase is added to a reaction vessel along with a surfactant-containing buffer and sample. After a short incubation, paramagnetic particles coated with monoclonal anti-cTnI antibody are added. The human cTnI binds to the anti-cTnI antibody on the solid phase, while the anti-cTnI antibody-alkaline phosphatase conjugate reacts with different antigenic sites on the cTnI molecules. After incubation in a reaction vessel, materials bound to the solid phase are held in a magnetic field while unbound materials are washed away. Then, the chemiluminescent substrate is added to the vessel and light generated by the reaction is measured with a luminometer. The light production is directly proportional to the concentration of cTnI in the sample. The amount of analyte in the sample is determined from a stored, multi-point calibration curve.(Package insert: ACCESS hsTnI. Beckman Coulter; 10/2021)

PDF Report
Indicates whether the report includes an additional document with charts, images or other enriched information

No

Day(s) Performed
Outlines the days the test is performed. This field reflects the day that the sample must be in the testing laboratory to begin the testing process and includes any specimen preparation and processing time before the test is performed. Some tests are listed as continuously performed, which means that assays are performed multiple times during the day.

Monday, Wednesday, Friday

Report Available
The interval of time (receipt of sample at Mayo Clinic Laboratories to results available) taking into account standard setup days and weekends. The first day is the time that it typically takes for a result to be available. The last day is the time it might take, accounting for any necessary repeated testing.

2 to 3 days

Specimen Retention Time
Outlines the length of time after testing that a specimen is kept in the laboratory before it is discarded

7 days

Performing Laboratory Location
Indicates the location of the laboratory that performs the test

Rochester

Fees
Several factors determine the fee charged to perform a test. Contact your U.S. or International Regional Manager for information about establishing a fee schedule or to learn more about resources to optimize test selection.

  • Authorized users can sign in to Test Prices for detailed fee information.
  • Clients without access to Test Prices can contact Customer Service 24 hours a day, seven days a week.
  • Prospective clients should contact their account representative. For assistance, contact Customer Service.

Test Classification
Provides information regarding the medical device classification for laboratory test kits and reagents. Tests may be classified as cleared or approved by the US Food and Drug Administration (FDA) and used per manufacturer instructions, or as products that do not undergo full FDA review and approval, and are then labeled as an Analyte Specific Reagent (ASR) product.

This test has been cleared, approved, or is exempt by the US Food and Drug Administration and is used per manufacturer's instructions. Performance characteristics were verified by Mayo Clinic in a manner consistent with CLIA requirements.

CPT Code Information
Provides guidance in determining the appropriate Current Procedural Terminology (CPT) code(s) information for each test or profile. The listed CPT codes reflect Mayo Clinic Laboratories interpretation of CPT coding requirements. It is the responsibility of each laboratory to determine correct CPT codes to use for billing.

CPT codes are provided by the performing laboratory.

84484

LOINC® Information
Provides guidance in determining the Logical Observation Identifiers Names and Codes (LOINC) values for the order and results codes of this test. LOINC values are provided by the performing laboratory.

Test Id Test Order Name Order LOINC Value
HSTNI Troponin I, High Sensitivity, P 89579-7
Result Id Test Result Name Result LOINC Value
Applies only to results expressed in units of measure originally reported by the performing laboratory. These values do not apply to results that are converted to other units of measure.
HSTNI Troponin I, High Sensitivity, P 89579-7

Test Setup Resources

Setup Files
Test setup information contains test file definition details to support order and result interfacing between Mayo Clinic Laboratories and your Laboratory Information System.

Excel | Pdf

Sample Reports
Normal and Abnormal sample reports are provided as references for report appearance.

Normal Reports | Abnormal Reports

SI Sample Reports
International System (SI) of Unit reports are provided for a limited number of tests. These reports are intended for international account use and are only available through MayoLINK accounts that have been defined to receive them.

SI Normal Reports | SI Abnormal Reports